How to calibrate a micrometer?
In addition to the common errors mentioned, other problems also occur in practice, for example due to insufficient calibration. Calibrating a micrometer is important to ensure measurement accuracy, especially in industrial environments or with tight manufacturing tolerances. We explain how to set calibration intervals in our blog about the types and characteristics of dial gauges.
You can also calibrate a micrometer yourself. A systematic procedure and a suited calibration standard are required for this.
Calibration involves checking whether the micrometer measures correctly within a permissible deviation. If the reading deviates from the specified reference measurement, the zero point can be readjusted. A suitable calibration standard, cleaning cloth and test record or metrology equipment card are required for calibration.
Step 1: Cleaning
Clean the lead screw, anvil, and standard calibration block with a soft, lint-free cloth. Only touch the micrometer by the handle to avoid heat transfer.
Step 2: Check the zero point
Slowly turn the lead screw until it stops against the anvil using the ratchet. The main scale must indicate exactly 0.00 mm, and the 0 drum mark must be aligned with the reference line. For many micrometers, the zero point can be readjusted. The special tool required for this purpose is normally supplied with the instrument.
Step 3: Insert and check the standard calibration block
Insert the standard calibration block between the anvil and lead screw and close it with the ratchet until it releases. Take the reading and compare it to the nominal dimension. Observe permissible tolerance limits.
Step 4: Repeat the test
Repeat several times to verify repeatability.
If the micrometer measures out of tolerance, the zero point may need to be readjusted, the measuring surfaces replaced or the measuring tool professionally recalibrated.
Self-calibration of a micrometer is easily possible with some care and a suitable reference dimension. However, it should be noted that such calibration is often not considered evidence in industrial quality processes. In such cases, the calibration must be performed with a traceable standard block and must be regularly confirmed by an accredited calibration laboratory.